Войти:
Оценок - 3, средний балл: 4 ( )

Инструкция по эксплуатации Fluke, модель IMPULSE 7000DP

Производитель: Fluke
Размер: 79.32 kb
Название файла: 7000dp__umeng0103.pdf
Язык инструкции:en
Перейти к скачиванию



Фрагмент инструкции


Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 % Lead I reference Performance waves and R wave detection Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 100 % 100 % 100 % Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 24 % 48 % 100 % 120 % 112 % 80 % ECG Normal Sinus Rates................................................................... 10 to 360 (by 1) BPM ECG High Level Output (BNC Jack) Amplitude Range ............................................................. 0.5 V per mV of reference lead setting Accuracy ......................................................... ±5 % Output Impedance .............................................. 50 . ECG on Defibrillator Input Load Same as the LEAD II amplitude but limited to ±4 mV ECG Performance Waves Square wave ....................................................... 2.0 and 0.125 Hz Triangular wave .................................................. 2.0 and 2.5 Hz 8/08 5 Manual Supplement Impulse 6000D/7000DP Users Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz Pulse................................................................... 30 and 60 BPM, 60 ms pulse width R-Wave Detection Waveform ........................................................... Haver-triangle Amplitude............................................................ 0.05 to 0.45 (by 0.05) mV 0.5 to 5.0 (by 0.5) mV Rate .................................................................... 30, 60, 80, 120, 200, and 250 BPM Widths................................................................. 8, 10, 12 ms, and 20 to 200 (by 10) ms Accuracy............................................................. ±(1 % setting + 1 ms) Noise Immunity Wave................................................................... Sine Line Frequency................................................... 50 or 60 Hz (± 0.5 Hz) Amplitude Range ............................................................. 0.0 to 10.0 (by 0.5) mV Accuracy......................................................... ±5 % Transvenous Pacer Pulse Simulation Widths Range ............................................................. 0.1, 0.2, 0.5, 1.0, and 2.0 ms Accuracy......................................................... ±5 % of setting Amplitude Range ............................................................. 0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and ±700 mV 6 4/08 Impulse 6000D/7000DP Users Manual Supplement Accuracy ......................................................... ±(10 % of setting + 0.2 mV) Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 67 % 100 % 33 % 67 % 67 % 67 % 67 % 67 % 67 % Lead I reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 100 % 100 % 100 % Arrhythmia Selections Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only) Demand .......................................................... 30 to 360 (by 1) BPM Asynchronous Non-Capture Non-Function Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA Supraventricular Atrial Fibrillation Coarse Atrial Fibrillation fine Atrial Flutter Sinus Arrhythmia Missed Beat Atrial Tachycardia 4/08 7 Manual Supplement Impulse 6000D/7000DP Users Change #2 On page 12, prior to Analyzing Pacemakers (7000DP only), add the following section: SCOPE OUTPUT Each time a defibrillator is fired into the Analyzer during an active defibrillator test, a delayed representation of the defibrillator pulse is sent to the Scope Output jack on the rear panel of the Analyzer. The Scope Output is an isolated signal and should not present any problems when connected to the input of an oscilloscope. To assist with the displaying of the defibrillator pulse, synchronizing marker pulses appear on the scope output jack along with the defibrillator pulse. Table 3A list these markers and describes their purpose. Table 3A. Synchronizing Marker Pulses Marker Pulse Characteristics Purpose Defib fire +2 V, 50 ms Indicates when the defibrillator was fired. Ranging -4 V, 0.4 ms, 1 ms apart Indicates the scaling used to output the pulse image: 1 pulse……….each volt out equals 80 volts pulse amplitude. 2 pulses……...each volt out equals 400 volts pulse amplitude. 3 pulses……...each volt out equals 2000 volts pulse amplitude. 8 4/08 Impulse 6000D/7000DP Users Manual Supplement Pulse playback +4 V, 1 ms pulse Start of defibrillator pulse playback [1] [1] Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early transitional signals. Use the synchronizing marker pulses to display specific information from the Scope Output signal. To see when the defibrillator fires, set the oscilloscope trigger level to +2 V. To get the ranging information, se...

Эта инструкция также подходит к моделям:
CD-проигрыватели - IMPULSE 6000D (79.32 kb)

Напишите ваш отзыв об устройстве



Текст отзыва
Ваше имя:
Введите две цифры:
capcha





Категории