Войти:
Оценок - 3, средний балл: 4 ( )

Инструкция по эксплуатации Invacare, модель 636959

Производитель: Invacare
Размер: 782.9 kb
Название файла: 636959.pdf
Язык инструкции:enesfr
Перейти к скачиванию



Фрагмент инструкции


6. Connect the suction tubing to the labeled patient connector (outlet) on the collection bottle. Part No 636959 3 Aspirator Suction Unit Elbow Connector DETAIL “A” Tubing Power Cord Vacuum Adjustment Knob Vacuum Gauge Collection Bottle Holder Hydrophobic Filter Connection Tubing Lid Power Connection Fuse On (I) Position Off (O) Position FIGURE 1 Medical Suction Unit Using the Suction Unit CAUTION Should fluid be aspirated back into the unit, send your suction unit to an authorized service center for a detailed inspection as the vacuum pump may have been damaged. NOTE: For this procedure, refer to FIGURE 1 and FIGURE 2 on page 5. NOTE: If the unit sustains a severe drop, validity of the gauge must be checked. Inspect if the float shut.off is still functioning properly when the liquid in the collection bottle reaches the safe full level. NOTE: Always transport unit with vacuum adjust knob rotated fully clockwise in case unit is dropped. 1. Ensure the suction tubing is connected to the labeled patient connector(outlet) on the collection bottle. 2. Ensure that all connections are secure and without leaks. 3. Connect the unit to power. 4. Turn the power switch On (I). 5. Adjust the vacuum level from 0 to 560 mmHg by turning the vacuum adjust knob: •Clockwise to increase vacuum. •Counter.clockwise to decrease vacuum. NOTE: Refer to the vacuum gauge while setting the desired level of vacuum. To accurately read the gauge, block the patient end of the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum reading. Aspirator Suction Unit 4 Part No 636959 6. Position the suction tubing appropriately with the patient and start therapy. NOTE: When the fluid reaches the safe full level of the collection bottle (800 cc), the suctioning will stop automatically (float shut.off). Refer to FIGURE 2. CAUTION Turn the power switch Off (O) when the automatic float shut-off is activated. Further suctioning could cause damage to the vacuum pump. The collection jar should be emptied after every use. NEVER let the fluid level exceed the “Safe Full” line. Overflow can damage the aspirator. 7. Turn the power switch Off (O) when the automatic float shut.off is activated. 8. Remove and clean the collection bottle when the motor has completely stopped and the vacuum has dropped. Refer to Cleaning the Collection Bottle and Lid. 9. Unplug the suction unit when not in use. Patient End of Suction Tubing Safe Full Level Collection Bottle FIGURE 2 Collection Bottle Components Cleaning the Collection Bottle and Lid NOTE: For this procedure, refer to FIGURE 1 on page 4 and FIGURE 2. NOTE: Clean the collection bottle and lid after each use. 1. Ensure the power switch is Off (O), the motor has completely stopped, and the vacuum has dropped. 2. Disconnect the suction unit from the AC power source. 3. Remove the elbow connector and tubing from the collection bottle lid. 4. Remove the collection bottle lid from the collection bottle. 5. Empty the contents. 6. Clean the collection bottle and lid after each use by one of the following methods: •Wash the collection jar and lid in the upper rack of a dishwasher with a water temperature maximum of 150oF (65oC). Part No 636959 5 Aspirator Suction Unit •Perform the following: (1) Wash with a hot water/dishwashing detergent solution; (2) Rinse with clean, hot tap water; (3) Soak in a one part vinegar to three parts hot water solution for 20 minutes; and (4) Rinse with hot tap water and air dry. •Wash with rubbing alcohol and air dry. •Wash with a commercial (bacterial.germicidal) disinfectant. Follow disinfectant manufacturer.s recommended instructions and dilution rates carefully. Cleaning the Suction Unit WARNING DO NOT submerge suction unit in water. This will result in severe damage to the vacuum pump. ALWAYS be sure to remove any foreign material (dirt, lint, etc.) from the suction unit. 1. Disconnect the suction unit from all external power sources. 2. Wipe the suction unit housing with a clean cloth and any commercial (bacterial.germicidal) disinfectant. Cleaning the Tubing 1. Disconnect the tubing from the connection jar. 2. Clean the tubing after each use by one of the following methods: •Wash with rubbing alcohol and air dry. •Perform the following (make sure all solutions run through the tubing): (1) Wash in a hot water/dishwashing detergent solution; (2) Rinse with clean, hot tap water; (3) Soak in one part vinegar to three parts hot water solution for 20 minutes; (4) Rinse with hot tap water; and, (5) Air dry. •Wash with a commercial (bacterial.germicidal) disinfectant. Follow disinfectant manufacturer.s recommended instructions and dilution rates carefully. 3. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth. Maintenance WARNING Dispose of collection bottle, tubing and hydrophobic filter in a specific waste bin labeled for medical waste. Residues in the collection bottle, tubing and filter may cause respiratory infection. Inspect tubin...


Напишите ваш отзыв об устройстве



Текст отзыва
Ваше имя:
Введите две цифры:
capcha





Категории